Single-Use Conductivity Sensors

Our PendoTECH Single-Use Conductivity Sensors measure raw conductivity and temperature with no calibration required because of pre-determined cell constant.

  • 4 – 20mA outputs for both conductivity and temperature
  • A pressure rating of 75 psi
  • Temperature normalization to 25°C
1140x660_53835266-9-min

Low Cost For Single-Use Applications

Our conductivity sensors are low cost, and have the option to add a panel mount kit, DIN rail kit, and benchtop stand to work in conjunction with the conductivity monitor.

1140x660_53835266-10-min

Optional One-Point Calibration By User

Our single-use conductivity sensors require no calibration because of pre-determined cell constant and the temperature measurement for compensation by our monitor is within a range from 0 to 70°C.

Overview

Features & Benefits

  • Performance with temperature correction
    • Range: 0.1 to 100 mS/cm
    • Accuracy: from 0.1 to 2 mS/cm +/- 0.1 mS/cm; 2 to 50 mS/cm +/- 5% of reading; 50 to 100 mS/cm typically +/- 5% of reading
    • Pressure Rating: 75 psi (5 bar)
  • Fluid path materials
    • Gamma compatible
    • NaOH resistant
    • USP Class VI
  • Conductivity monitor details:
    • Ability to read 2 sensors; including temperature only
    • Performs temperature normalization to 25°C
    • Analog 4 - 20mA outputs for conductivity and temperature to enable integration to a control system
    • RS232 output for data collection to a PC

Applications

  • Diafiltration
  • Tangential Flow Filtration (TFF)
  • Virus Filtration
  • Harvesting
  • Fill and Finish
  • Perfusion
  • Nano Filtration
  • Chromatography
  • Sterile Filtration
  • Research
  • Process Development
  • As well as many other applications

Technical Specs

Accuracy From 0.1 to 2 mS/cm +/- 0.1 mS/cm; 2 to 50 mS/cm +/- 5% of reading; 50 to 100 mS/cm typically +/- 5% of reading
Pressure Range 75 psi max
Biocompatibility All materials in contact with product fluid path meet USP Class VI requirements, both pre and post gamma exposure
Gamma Irradiation Up to 50 kiloGrays
Operating Temperature 2° C to 50° C (other ranges with process qualification) thermistor reads to 70° C
ADCF Status All fluid path materials are animal derived component free
Temperature Range 0 to 70 °C
Temperature Accuracy Better than +/- 0.2°C (typical better than 0.1°C)
Temperature Element Thermistor with resistance @25°C of 2252 ohm
Connector Custom molded water-tight 4 pin connector
Packaging Sealed in vapor barrier bag inside polybag