Single-Use Conductivity Sensors
Low Cost For Single-Use Applications
Our conductivity sensors are low cost, and have the option to add a panel mount kit, DIN rail kit, and benchtop stand to work in conjunction with the conductivity monitor.
Optional One-Point Calibration By User
Our single-use conductivity sensors require no calibration because of pre-determined cell constant and the temperature measurement for compensation by our monitor is within a range from 0 to 70°C.
Overview
PendoTECH Single-Use Conductivity Sensors and Conductivity monitor integrate for highly accurate conductivity and temperature measurement without the need to calibrate the conductivity electrode because of its predetermined cell constant. For use in critical bioprocess operations, the sensor monitor reads conductivity and temperature for up to two sensors. There is temperature normalization to 25°C with an adjustable mS/ °C factor. The monitor has 4 – 20mA outputs for both conductivity and temperature and an RS232 output for data collection to a PC. With a pressure rating of 75 psi, the PendoTECH Single-Use Conductivity Sensors materials that are in contact with the product fluid path meet USP Class VI requirements, both pre and post gamma exposure.
Features & Benefits
- Performance with temperature correction
- Range: 0.1 to 100 mS/cm
- Accuracy: from 0.1 to 2 mS/cm +/- 0.1 mS/cm; 2 to 50 mS/cm +/- 5% of reading; 50 to 100 mS/cm typically +/- 5% of reading
- Pressure Rating: 75 psi (5 bar)
- Fluid path materials
- Gamma compatible
- NaOH resistant
- USP Class VI
- Conductivity monitor details:
- Ability to read 2 sensors; including temperature only
- Performs temperature normalization to 25°C
- Analog 4 - 20mA outputs for conductivity and temperature to enable integration to a control system
- RS232 output for data collection to a PC
Documents
Applications
- Diafiltration
- Tangential Flow Filtration (TFF)
- Virus Filtration
- Harvesting
- Fill and Finish
- Perfusion
- Nano Filtration
- Chromatography
- Sterile Filtration
- Research
- Process Development
- As well as many other applications
Technical Specs
Accuracy | From 0.1 to 2 mS/cm +/- 0.1 mS/cm; 2 to 50 mS/cm +/- 5% of reading; 50 to 100 mS/cm typically +/- 5% of reading |
Pressure Range | 75 psi max |
Biocompatibility | All materials in contact with product fluid path meet USP Class VI requirements, both pre and post gamma exposure |
Gamma Irradiation | Up to 50 kiloGrays |
Operating Temperature | 2° C to 50° C (other ranges with process qualification) thermistor reads to 70° C |
ADCF Status | All fluid path materials are animal derived component free |
Temperature Range | 0 to 70 °C |
Temperature Accuracy | Better than +/- 0.2°C (typical better than 0.1°C) |
Temperature Element | Thermistor with resistance @25°C of 2252 ohm |
Connector | Custom molded water-tight 4 pin connector |
Packaging | Sealed in vapor barrier bag inside polybag |